Design and performance characteristics of the radiation shield developed for the TFTR charge exchange diagnostics
Creators
- 1. Princeton University Plasma Physics Laboratory, Princeton, New Jersey 08543
Description
The high source strength of 2.45-MeV neutrons (∼1.2 x 10/sup 16/ n/s) and neutron-induced gamma rays produced by the Tokamak Fusion Test Reactor (TFTR) have substantially degraded the performance of many diagnostics. In particular, data obtained by the charge exchange neutral analyzer (CENA) were seen to degrade at modest production levels of ∼10/sup 15/ n/s. To reduce the radiation-induced noise to acceptable levels, massive shielding was required. A design and cost-effectiveness study was performed for the TFTR CENA which showed that a double-layered enclosure consisting of a 10-cm-thick inner housing of lead surrounded by a 23-cm-thick layer of 1.0% borated polyethylene was required to reduce the neutron and gamma noise level by two orders of magnitude. This design was recently implemented on TFTR and was shown to perform as predicted. The details of the enclosure will be presented and the contributing factors that led to the choice of this particular design will be reviewed
Additional details
Publishing Information
- Journal Title
- Rev. Sci. Instrum.
- Journal Volume
- 59
- Journal Issue
- 8
- Series
- Rev. Sci. Instrum.
- Journal Page Range
- 1726-1728
- ISSN
- 0034-6748
- CODEN
- RSINA
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 19097096
- Subject category
- S70: PLASMA PHYSICS AND FUSION TECHNOLOGY; S70: PLASMA PHYSICS AND FUSION TECHNOLOGY;
- Descriptors DEI
- ELECTRON MULTIPLIER DETECTORS; NEUTRON DETECTORS; PERFORMANCE; PLASMA DIAGNOSTICS; RADIATION PROTECTION; SHIELDS; TFTR TOKAMAK
- Descriptors DEC
- CLOSED PLASMA DEVICES; MEASURING INSTRUMENTS; RADIATION DETECTORS; THERMONUCLEAR DEVICES; TOKAMAK DEVICES