On the accuracy of localization achievable in fiducial-based stereoscopic image registration system using an electronic portal imaging device
Creators
- 1. Clinical Oncology Unit, Faculty of Medicine, University of Malaya, Kuala Lumpur, (Malaysia)
- 2. School of Physics, The University of Western Australia, Crawley, WA (Australia)
Description
Portal imaging using electronic portal imaging device (EPID) is a well-established image-guided radiation therapy (IGRT) technique for external beam radiation therapy. The aims of this study are threefold; (i) to assess the accuracy of isocentre localization in the fiducial-based stereoscopic image registration, (ii) to investigate the impact of errors in the beam collimation device on stereoscopic registration, and (iii) to evaluate the intra- and inter-observer variability in stereoscopic registration. Portal images of a ball bearing phantom were acquired and stereoscopic image registrations were performed based on a point centred in the ball bearing as the surrogate for registration. Experiments were replicated by applying intentional offsets in the beam collimation device to simulate collimation errors. The accuracy of fiducial markers localization was performed by repeating the experiment using three spherical lead shots implanted in a pelvic phantom. Portal images of pelvis phantom were given to four expert users to assess the inter-observer variability in performing registration. The isocentre localization accuracy tested using ball bearing phantom was within 0.3 mm. Gravity-induced systematic errors of beam collimation device by 2 mm resulted in positioning offsets of the order of 2 mm opposing the simulated errors. Relatively large inter-portal pair projection errors ranges from 1.3 mm to 1.8 mm were observed with simulated errors in the beam collimation device. The intra-user and inter-user variabilities were observed to be 0.8 and 0.4 mm respectively. Fiducial-based stereoscopic image registration using EPID is robust for IGRT procedure.
Availability note (English)
Available from http://rd.springer.com/article/10.1007/s13246-012-0148-yAdditional details
Identifiers
Publishing Information
- Journal Title
- Australasian Physical and Engineering Sciences in Medicine
- Journal Volume
- 35
- Journal Issue
- 2
- Journal Page Range
- p. 205-213
- ISSN
- 0158-9938
INIS
- Country of Publication
- Australia
- Country of Input or Organization
- Australia
- INIS RN
- 46037710
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCURACY; BEAM PROFILES; BIOLOGICAL LOCALIZATION; COLLIMATORS; EVALUATION; EXTERNAL BEAM RADIATION THERAPY; FOCUSING; IMAGE PROCESSING; PATIENTS; PHANTOMS; RADIATION SOURCE IMPLANTS; RANDOMNESS
- Descriptors DEC
- IMPLANTS; MEDICINE; MOCKUP; NUCLEAR MEDICINE; PROCESSING; RADIATION SOURCES; RADIOLOGY; RADIOTHERAPY; STRUCTURAL MODELS; THERAPY
- Proposed descriptors and Free-text terms
- FIDUCIAL MARKERS
Optional Information
- Notes
- 5 figs., 1 tab.