Published December 11, 2012 | Version v1
Journal article

Characterization and Optimization of silicon photomultipliers for the development of intraoperative beta probes

  • 1. Laboratoire Imagerie et Modélisation en Neurobiologie et Cancérologie (UMR 8165), Campus d'Orsay, IN2P3-CNRS, 91406 Orsay Cedex (France)
  • 2. Laboratoire de l'Accélérateur Linéaire, IN2P3-CNRS, 91898 Orsay (France)

Description

This work reports the results of a comprehensive study aiming to characterize and optimize the performances of Silicon Photomultipliers (SiPM) as photodetectors for intraoperative beta detection of radiolabeled tumor tissues. In that purpose, we focused our measurements on the influence of temperature and operating voltage on dark noise (count rate and amplitude distribution), absolute photon detection efficiency and total gain of different SiPM devices from Hamamatsu. The impact of these parameters on the overall beta sensitivity of SiPM-based scintillation detectors was quantified as a function of the intensity of the incident light following a simple physical model that is shown to correspond well with measurements.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2011.10.046

Additional details

Identifiers

DOI
10.1016/j.nima.2011.10.046;
PII
S0168-9002(11)01996-6;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
695
Journal Page Range
p. 242-246
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
6. international conference on new developments in protodetection
Acronym
NDIP11
Dates
4-8 Jul 2011
Place
Lyon (France)

Optional Information

Copyright