Quality Assurance of Radiation Therapy Planning Systems: Current Status and Remaining Challenges
Creators
- 1. Departments of Oncology, Medical Biophysics, Diagnostic Imaging and Nuclear Medicine, and Physics and Astronomy, University of Western Ontario, London, ON (Canada)
- 2. Department of Physics and Engineering, London Regional Cancer Program/London Health Sciences Centre, London, ON (Canada)
Description
Computerized radiation therapy planning systems (RTPSs) are pivotal for treatment planning. The acceptance, commissioning, and quality control of RTPSs are uniquely complex and are described in the American Association of Physicists in Medicine Task Group Report 53 (1998) and International Atomic Energy Agency Technical Report Series No. 430 (2004). The International Atomic Energy Agency also developed a document and data package for use by vendors and purchasers to aid with acceptance testing of RTPSs. This document is based on International Electrotechnical Commission standard 62083 (2000) and describes both 'type' tests to be performed in the factory and 'site' tests to be performed in the clinic. The American Association of Physicists Task Group Report 67 described benchmark tests for the validation of dose calculation algorithms. Test data are being produced with the backing of the U.S. National Cancer Institute. However, significant challenges remain. Technology keeps evolving rapidly, thus requiring new quality assurance (QA) procedures. Intensity-modulated radiation therapy with its use of inverse optimization has added a new dimension to QA, because the results are not intuitively obvious. New technologies such as real-time ultrasound guidance for brachytherapy, TomoTherapy, and Cyberknife, require their own specialized RTPSs with unique QA requirements. On-line imaging allows for the generation of dose reconstructions using image warping techniques to determine the daily dose delivered to the patient. With increasing computer speeds, real-time reoptimization of treatment plans will become a reality. Gating technologies will require four-dimensional dose calculations to determine the actual dose delivered to tissue voxels. With these rapidly changing technologies, it is essential that a strong QA culture is invoked in every institution implementing these procedures and that new protocols are developed as a part of the clinical implementation process
Availability note (English)
Available from http://dx.doi.org/10.1016/j.ijrobp.2007.04.095Additional details
Identifiers
- DOI
- 10.1016/j.ijrobp.2007.04.095;
- PII
- S0360-3016(07)04273-3;
Publishing Information
- Journal Title
- International Journal of Radiation Oncology, Biology and Physics
- Journal Volume
- 71
- Journal Issue
- 1
- Journal Page Range
- p. S23-S27
- ISSN
- 0360-3016
- CODEN
- IOBPD3
Conference
- Title
- Challenges of advanced technology
- Acronym
- 2007 interorganizational symposium on quality assurance of radiation therapy
- Dates
- 20-22 Feb 2007
- Place
- Dallas, TX (United States)
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 40006922
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ALGORITHMS; BRACHYTHERAPY; COMPUTERIZED TOMOGRAPHY; CT-GUIDED RADIOTHERAPY; DRUGS; IMAGES; INTERNATIONAL ELECTROTECHNICAL COMMISSION; NEOPLASMS; OPTIMIZATION; PATIENTS; PLANNING; QUALITY ASSURANCE; QUALITY CONTROL; RADIATION DOSES; VALIDATION
- Descriptors DEC
- CONTROL; DIAGNOSTIC TECHNIQUES; DISEASES; DOSES; INTERNATIONAL ORGANIZATIONS; MATHEMATICAL LOGIC; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; RADIOTHERAPY; TESTING; THERAPY; TOMOGRAPHY
Optional Information
- Copyright
- Copyright (c) 2008 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.