Published September 1979 | Version v1
Journal article

A system for time-resolved x-ray diffraction and its application to muscle contraction

  • 1. National Lab. for High Energy Physics, Oho, Ibaraki (Japan)

Description

A data-collection system has been built which permits time-resolved studies of X-ray diffraction diagrams obtained from contracting muscle on millisecond time scale. The system consists of a linear delay-line position sensitive proportional counter (PSPC), a special data transfer unit and an on-line computer. The PSPC used with a mirror-monochromator camera can detect equatorial reflections from stimulated muscle in a total exposure time of a few seconds. Time-resolved data-collection is achieved by stimulating muscle at a regular time interval, dividing a complete cycle of muscle contraction into many successive time slices and accumulating in computer memory X-ray data for each time slice from many repeated cycles of stimulation. The performances of the system have been demonstrated by recording equatorial reflections from frog skeletal muscle during isometric and isotonic twitch with a time resolution of 25 ms. (author)

Additional details

Publishing Information

Journal Title
J. Fac. Eng., Univ. Tokyo, Ser. B
Journal Volume
35
Journal Issue
2
Series
J. Fac. Eng., Univ. Tokyo, Ser. B.
Journal Page Range
313-324
ISSN
0563-7937