Published December 1, 2012 | Version v1
Journal article

A diffusion chamber for passive separated measurements of radon/thoron concentration in dwellings

  • 1. Faculty of Nuclear engineering and physics, Amirkabir University of Technology, Tehran (Iran, Islamic Republic of)
  • 2. Radiation Application Research School, Nuclear Science and Technology Research Institute, Tehran (Iran, Islamic Republic of)
  • 3. Iran Nuclear Regulatory Authority, Karegar St., P.O. Box 14155-1339, Tehran (Iran, Islamic Republic of)

Description

In this research, a passive diffusion chamber has been developed for separated measurement of radon and thoron. The chamber consists of two volumes which are separated by a fiber glass filter. Two lexan polycarbonate nuclear track detectors (film) are placed inside of the volumes to detect the alpha particles of radon/thorn and/or their progenies. Another lexan polycarbonate detector covered with an optimized thickness of an attenuator is placed outside of the chamber to measure only 212Po which its concentration can be related to that of long-life thoron progeny, 212Pb. The sensitivities have been measured by 2.06 and 0.053 [tracks cm−2(kBq m−3 day)−1] values for radon and thoron respectively inside of the chamber, and 7960 [tracks cm−2(kBq m−3 day)−1] value for thoron outside of the chamber. So the system can be successfully used for separated measurement of an extended range of radon/thoron concentrations in dwellings.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2012.08.080

Additional details

Identifiers

DOI
10.1016/j.nima.2012.08.080;
PII
S0168-9002(12)00983-7;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
694
Journal Page Range
p. 331-334
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright