Published October 2016
| Version v1
Journal article
Design and Optimization of an Analog Filter with a CdTe Detector for X-ray Fluorescence Applications
- 1. Sungkyunkwan University, Suwon (Korea, Republic of)
- 2. Korea Atomic Energy Research Institute, Jeongeup (Korea, Republic of)
Description
An analog pre-filter circuit for digital pulse processing is designed and optimized for X-ray fluorescence (XRF) applications to replace traditional analog shaping amplifiers. To optimize the pre-filter performance, we characterized noise electrons as a function of the input pulse rise time and decay time of the output pulse by using the full width at half maximum. In addition, gamma-ray energy measurements at room temperature showed that the commercially available CdTe Schottky-type radiation detector with our newly designed and optimized pre-filter circuit exhibited full widths at half maxima of 4.97 (Ba-133, at 53 keV) and 5.56 keV (Am-241, at 59.5 keV), respectively.
Additional details
Publishing Information
- Journal Title
- Journal of the Korean Physical Society
- Journal Volume
- 69
- Journal Issue
- 8
- Series
- 15 refs, 7 figs, 2 tabs
- Journal Page Range
- p. 1157-1162
- ISSN
- 0374-4884
INIS
- Country of Publication
- Korea, Republic of
- Country of Input or Organization
- Korea, Republic of
- INIS RN
- 49096429
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- AMPLIFIERS; CDTE SEMICONDUCTOR DETECTORS; DESIGN; FLUORESCENCE; GAMMA RADIATION; OPTIMIZATION; PERFORMANCE; RADIATION DETECTORS; SCHOTTKY DEFECTS; USES
- Descriptors DEC
- CRYSTAL DEFECTS; CRYSTAL STRUCTURE; ELECTROMAGNETIC RADIATION; ELECTRONIC EQUIPMENT; EMISSION; EQUIPMENT; IONIZING RADIATIONS; LUMINESCENCE; MEASURING INSTRUMENTS; PHOTON EMISSION; POINT DEFECTS; RADIATION DETECTORS; RADIATIONS; SEMICONDUCTOR DETECTORS; VACANCIES