Published October 2016 | Version v1
Journal article

Design and Optimization of an Analog Filter with a CdTe Detector for X-ray Fluorescence Applications

  • 1. Sungkyunkwan University, Suwon (Korea, Republic of)
  • 2. Korea Atomic Energy Research Institute, Jeongeup (Korea, Republic of)

Description

An analog pre-filter circuit for digital pulse processing is designed and optimized for X-ray fluorescence (XRF) applications to replace traditional analog shaping amplifiers. To optimize the pre-filter performance, we characterized noise electrons as a function of the input pulse rise time and decay time of the output pulse by using the full width at half maximum. In addition, gamma-ray energy measurements at room temperature showed that the commercially available CdTe Schottky-type radiation detector with our newly designed and optimized pre-filter circuit exhibited full widths at half maxima of 4.97 (Ba-133, at 53 keV) and 5.56 keV (Am-241, at 59.5 keV), respectively.

Additional details

Publishing Information

Journal Title
Journal of the Korean Physical Society
Journal Volume
69
Journal Issue
8
Series
15 refs, 7 figs, 2 tabs
Journal Page Range
p. 1157-1162
ISSN