Dosimetric properties of brass mesh bolus for high-energy photon beam

  • 1. Kyushu University Hospital, Department of Medical Technology, Division of Radiology, Fukuoka (Japan)

Description

To investigate fundamental dosimetric properties of surface dose, exit dose, and beam profile of the brass mesh bolus for 4, 6, and 10 MV high-energy photon beams in radiation therapy. Surface dose and exit dose in the water-equivalent phantom were measured, and percent depth doses (PDDs) were calculated with no bolus, one layer of brass mesh, two layers of brass mesh bolus, three layers of brass mesh bolus, and 0.5 cm tissue-equivalent (TE) bolus. Exit dose was measured at a phantom thickness of 10 cm. Beam profiles were measured at phantom depths of 0 cm and 10 cm. All dosimetry was performed for 4, 6, and 10 MV photon beams using a linear accelerator. The surface dose at a phantom depth of 0 cm increased to 37.3%, 36,3%, and 31.0% for 4, 6, and 10 MV, respectively, with the brass mesh bolus compared to the case of no bolus. The surface dose decreased with one layer of brass mesh bolus compared to that with the 0.5 cm TE bolus. On the other hand, the exit dose increased to 22,0%, 23.1%, and 22.8% for 4, 6, and 10 MV, respectively, with the brass mesh bolus compared to the case of no bolus. The beam profile at the depth of 0 cm showed oscillations, and the difference between the maximum and minimum doses was up to 13,1% with one layer of brass mesh bolus. It was suggested that the brass mesh bolus not only increases the surface dose but also has different properties from the conventional TE bolus. (author)

Availability note (English)

Available from DOI: https://doi.org/10.6009/jjrt.2023-1323

Abstract (Japanese)

4, 6, 10 MV高エネルギー光子線に対する真ちゅう製ボーラスの表面線量,射出線量およびビームプロファイルの基礎特性を検証すること.真ちゅう製ボーラスと組織等価ボーラスを使用し,4, 6, 10 MV光子線の水等価ファントム中における表面線量とファントム厚さ10 cmにおける射出線量を測定しPDDを算出した.また,ファントム深さ0 cmと10 cmにおけるビームプロファイルを取得した.真ちゅう製ボーラスありの場合,ボーラスなしの場合と比較して表面線量は4, 6, 10 MVにおいてそれぞれ37.3%, 36.3%, 31.0%増加したが,組織等価ボーラスと比較すると真ちゅう製ボーラス1枚での表面線量は減少した.射出線量は真ちゅう製ボーラスありの場合,ボーラスなしの場合と比較して4, 6, 10 MVにおいてそれぞれ22.0%, 23.1%, 22.8%増加した.真ちゅう製ボーラスは組織等価ボーラスとは異なる特性をもつことが示された.(著者)

Additional details

Additional titles

Original title (Japanese)
高エネルギー光子線に対する真ちゅう製ボーラスの線量特性