Published January 2008 | Version v1
Journal article

Dosimetric Effects of Setup Uncertainties on Breast Treatment Delivery

  • 1. Regional Medical Physics Department and Northern Centre for Cancer Treatment, Newcastle General Hospital, Newcastle Upon Tyne (United Kingdom)

Description

This study aimed to assess the dosimetric impact of setup errors during the delivery of radiotherapy to the breast, and use this information to make recommendations on intervention tolerances for portal imaging of breast treatments. Translational and rotational setup errors were simulated for 10 recent breast patients using an Oncentra MasterPlan treatment planning system. The effect of these errors on the breast and tumor bed target volumes receiving 95% and 107% of the prescribed dose were assessed. For the majority of patients, shifts of up to 10 mm or a 4 deg. patient rotation about the cranio-caudal axis had no significant effect on the dose distribution. Changes in dosimetry were more likely if the reference plan contained large hot or cold spots. For a typical patient, it is estimated that a shift of 5 mm in any one direction, or a 2 deg. patient rotation would not cause more than a 5% change in the target volume receiving between 95% and 107% of the prescribed dose. If combinations of errors occur, greater dosimetric changes would be expected. It is concluded that individual patient shifts of up to 5 mm or rotations about the cranio-caudal axis of 2 deg. or less are unlikely to affect dose-volume histogram parameters by an amount judged as clinically significant. Setup errors exceeding these values may cause large dosimetric changes for some patients, particularly those with larger hot or cold regions in the dose distribution, and intervention is therefore recommended

Availability note (English)

Available from http://dx.doi.org/10.1016/j.meddos.2008.01.003

Additional details

Identifiers

DOI
10.1016/j.meddos.2008.01.003;
PII
S0958-3947(08)00035-6;

Publishing Information

Journal Title
Medical Dosimetry
Journal Volume
33
Journal Issue
4
Journal Page Range
p. 293-298
ISSN
0958-3947

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
40054631
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
CARCINOMAS; DOSIMETRY; ERRORS; MAMMARY GLANDS; PATIENTS; PLANNING; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; RADIOTHERAPY; RECOMMENDATIONS; SIMULATION; TOLERANCE
Descriptors DEC
BODY; DISEASES; DOSES; GLANDS; MEDICINE; NEOPLASMS; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; THERAPY