Published April 2006 | Version v1
Journal article

A comparative study in three commercial kits of radioligand assay for GAD-Ab

  • 1. Institute of Metabolism and Endocrinology, Second Xiangya Hospital, Central South Univ., The Diabetic Center of Central South Univ., Changsha (China)

Description

Objective: To provide data for perfect selection of glutamic acid decarboxylase antibodies (GAD-Ab) assay commercial kits. Methods: The sera of 88 titer-graded samples by radioligand assay (RLA) were measured with two types of enzyme-linked immunosorbent assay (ELISA) kits and one type of immunoradiometric assay (IRMA) kit for GAD-Ab in order to test their consistency with RLA. Another 140 diabetic sera were measured with RLA and screened by a type of suitable kit to search the cutoff point for discriminating patients who needed rechecking. Results: Compared with RLA, concordance ratio of 3 kits: Medizym kit(75%) > RSR kit(73.9%) > Biomerica kit(62.5%); Kappa value: Medizym kit (0.408, common) > RSR kit (0.405, common) > Biomerica kit (0.185, failing); correlation to RLA index: RSR kit (r= 0.992) > Medizym kit(r= 0. 791 ) > Biomerica kit (r = -0. 055); area under the receiver operating characteristic (ROC) curve: Medizym kit (0.812) > RSR kit (0. 727 ) > Biomerica kit (0.666). Conclusions: The efficiencies of RSR kit and Medizym kit are good. Serum concentration between 0.25-0.99 U/ml measured by RSR kit is suggested to further recheck by RLA. (authors)

Additional details

Publishing Information

Journal Title
Chinese Journal of Nuclear Medicine
Journal Volume
26
Journal Issue
2
Journal Page Range
p. 100-103
ISSN
0253-9780

Optional Information

Notes
3 tabs., 10 refs.