Design and fabrication of endoscope-type Compton camera
Creators
- 1. Department of Bio Engineering, The University of Tokyo (Japan)
- 2. Department of Nuclear Engineering and Management, The University of Tokyo (Japan)
Description
We are constructing an endoscope coupled with a Compton camera to intra-operatively inspect early stage cancer and metastatic lymph node (5 mm–10 mm). The radiation imaging system is composed of pixelated semiconductor detectors, which are Si array and CdTe array, whose sizes are less than 10 mm×10 mm, and a digital signal-processing unit with ASIC and FPGA and reconstruction algorithm using spherical harmonics that can compute in real time. In this paper, we show some simulation results on the performance of the proposed prototype detector using EGS5, Monte Carlo simulation code. The FWHM of 1 mm spatial resolution for an object located 10 mm away from the detector surface and the intrinsic efficiency of 0.05% were observed. Imaging and data acquisition time to take fine images are just 1 min. It therefore can be considered that the endoscopic Compton camera is useful for intra-operative inspection
Availability note (English)
Available from http://dx.doi.org/10.1016/j.nima.2013.06.100Additional details
Identifiers
- DOI
- 10.1016/j.nima.2013.06.100;
- PII
- S0168-9002(13)00954-6;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 731
- Journal Page Range
- p. 283-287
- ISSN
- 0168-9002
- CODEN
- NIMAER
Conference
- Title
- 6. international workshop on semiconductor pixel detectors for particles and imaging
- Acronym
- PIXEL 2012
- Dates
- 3-7 Sep 2012
- Place
- Inawashiro, Fukushima (Japan)
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 45051578
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ALGORITHMS; CAMERAS; CDTE SEMICONDUCTOR DETECTORS; COMPUTERIZED SIMULATION; DATA ACQUISITION; DESIGN; DIAGNOSTIC USES; EFFICIENCY; FABRICATION; IMAGE PROCESSING; MONTE CARLO METHOD; POSITION SENSITIVE DETECTORS; SI SEMICONDUCTOR DETECTORS; SPATIAL RESOLUTION; SPHERICAL HARMONICS
- Descriptors DEC
- CALCULATION METHODS; FUNCTIONS; MATHEMATICAL LOGIC; MEASURING INSTRUMENTS; PROCESSING; RADIATION DETECTORS; RESOLUTION; SEMICONDUCTOR DETECTORS; SIMULATION; USES
Optional Information