Published March 1, 2000 | Version v1
Journal article

Accurate EPR radiosensitivity calibration using small sample masses

Description

We demonstrate a procedure in retrospective EPR dosimetry which allows for virtually nondestructive sample evaluation in terms of sample irradiations. For this procedure to work, it is shown that corrections must be made for cavity response characteristics when using variable mass samples. Likewise, methods are employed to correct for empty tube signals, sample anisotropy and frequency drift while considering the effects of dose distribution optimization. A demonstration of the method's utility is given by comparing sample portions evaluated using both the described methodology and standard full sample additive dose techniques. The samples used in this study are tooth enamel from teeth removed during routine dental care. We show that by making all the recommended corrections, very small masses can be both accurately measured and correlated with measurements of other samples. Some issues relating to dose distribution optimization are also addressed

Additional details

Identifiers

PII
S0168900299009766;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
441
Journal Issue
3
Journal Page Range
p. 535-550
ISSN
0168-9002
CODEN
NIMAER

INIS

Country of Publication
Netherlands
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
34043072
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
CALIBRATION; DOSIMETRY; EPR SPECTROMETERS; NONDESTRUCTIVE TESTING; RADIATION DOSE DISTRIBUTIONS; TEETH
Descriptors DEC
DIGESTIVE SYSTEM; MATERIALS TESTING; MEASURING INSTRUMENTS; ORAL CAVITY; SPECTROMETERS; TESTING

Optional Information

Copyright
Copyright (c) 2000 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.