SU-E-T-97: Dependence Of Optically Stimulated Luminescent Dosimeter (OSLD) Out Of Field Response On Volumetric Modulated Arc Therapy (VMAT) Field Modulation

  • 1. Mayo Clinic Arizona, Phoenix, AZ (United States)

Description

Purpose: To determine the out of field response of Microstar ii OSLDs as a function of field modulation and distance in VMAT plan delivery. This work has potential application in fetal dose monitoring or measurements on cardiac pacemakers Methods: VMAT plans were created in Eclipse and optimized to varying degrees of modulation. Three plans were chosen to represent low, medium and high degrees of modulation (modulation factors as defined by MU/cGy). Plans were delivered to slabs of solid water with dimensions 60cm length, 30cm width, and 10cm height. For each modulation factor, 2 OSLDs were placed at 1cm depth with out of field distances of 1, 2, 3, 5, 8 and 10cm and the plan delivered isocentrically to a depth of 5cm. This technique was repeated for a Farmer Chamber by incrementing the table by the appropriate distance. The charge readings for the Farmer Chamber were converted to dose and the ratios taken as functions of modulation factors and distances out of field Results: Examination of the results as a function of out of field distance shows a trend of increasing OSLD/Farmer Chamber ratios for all modulation factors. The slopes appear to be roughly equivalent for all modulation factors investigated. Results as a function of modulation showed a downward trend for all out of field distances, with the greatest differences seen at 5cm and 8cm Conclusion: This study demonstrates that the response of OSLD dosimeters change as a function of out of field distance and modulation. The differences seen are within the stated accuracy of the system for the out of field distances and modulations investigated. Additional investigation is warranted to see if the OSLD response changes appreciably with longer out of field distances or wider ranges of modulation

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
41
Journal Issue
6
Journal Page Range
p. 244
ISSN
0094-2405
CODEN
MPHYA6

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
46046287
Subject category
S60: APPLIED LIFE SCIENCES;
Descriptors DEI
ACCURACY; CARDIAC PACEMAKERS; DOSEMETERS; LUMINESCENCE; MODULATION; RADIATION DOSES; RADIOTHERAPY
Descriptors DEC
DOSES; EMISSION; MEASURING INSTRUMENTS; MEDICINE; NUCLEAR MEDICINE; PHOTON EMISSION; RADIOLOGY; THERAPY

Optional Information

Notes
(c) 2014 American Association of Physicists in Medicine