Calculation of the systematic error and correction factors in gamma waste assay system

  • 1. Technical University of Budapest (Hungary). Inst. of Nuclear Techniques

Description

A model for the calculation of systematic errors and correction factors in a Segmented Gamma Scanner is given by simulating the measurement system. The calculational model was verified by experiment. The calculation values exhibit good agreement with experimental values. Through a simple code the results are used for estimating the range of measured uncertainty, and for giving the correction terms to improve accuracy. (Author)