Published 1994 | Version v1
Miscellaneous

Air filtration system for monitoring surface air pollution by man-made radionuclides

Description

A filtration device was constructed allowing effective sampling of surface air for further gamma spectrometric analysis. The working surface of the device is 0.5-1 m2, the capacity being 1800-4400 m3 per hour ±3.6%. The features of the device are illustrated by gamma spectrum of an aerosol sample obtained by using a spectrometer with Ge(Li) detector. Activities of 10-5 Bq.m-3 are determined with high degree of confidence. The very well pronounced pick of Be-7 could be considered as a measure of a high sensitivity and effectiveness of the air-filtration system. The cylindrical geometry of the measurements contributes also for the small value of errors. 30 minutes up to 1 hour is the time needed to get a sample with activity sufficient for obtaining reliable data. Results for daily variations of Be-7 concentration as well as for activities of Cs-134 and Cs-137 are presented. 3 tabs., 1 fig., 7 refs

Availability note (English)

Available from the Bulgarian INIS Centre, 69 Shipchenski Prokhod St., 1574 Sofia (BG).

Additional details

Additional titles

Original title (Bulgarian)
Vyzdukhofiltrirashta sistema za opredelyane zamyrsenostta na prizemniya vyzdushen sloj s tekhnogenni radionuklidi

Publishing Information

Imprint Pagination
10 p.

Conference

Title
National seminar with international participation 'Environmental radioactivity'.
Dates
11-13 Oct 1994.
Place
Sofia (Bulgaria).