Published April 1989 | Version v1
Journal article

Josephson array voltage calibration system

  • 1. National Inst. of Standards and Technology, Gaithersburg, MD (USA)

Description

The authors describe a new Josephson array system which maintains the U.S. Legal Volt. This system is almost fully automated, operates with a typical precision of 0.009 μV, and readily allows U.S. Legal Volt measurements weekly, or more frequently if desired. This system was compared to the previous volt maintenance system, and agreement was made to within 0.03 ppm. This verification is limited by uncertainties in the resistive divider instruments of the previous system

Additional details

Additional titles

Subtitle (English)
Operational use and verification

Publishing Information

Journal Title
IEEE Transactions on Instrumentation and Measurement
Journal Volume
38
Journal Issue
2
Series
IEEE Trans. Instrum. Meas.
Journal Page Range
296-301
ISSN
0018-9456
CODEN
IEIMA