Published December 2021 | Version v1
Journal article

Evaluation of a sensitive, reasonable, and fast detection method for 55Fe in steel

  • 1. Helmholtz-Zentrum Dresden-Rossendorf, Dresden (Germany)
  • 2. University of Vienna, Vienna (Austria). Faculty of Physics, Isotope Physics, VERA Laboratory
  • 3. VKTA - Strahlenschutz, Analytik & Entsorgung Rossendorf e. V., Dresden (Germany)

Description

A pilot study to quantify 55Fe in steel from a reactor vessel of a nuclear power plant by accelerator mass spectrometry (AMS) without any chemical sample preparation was validated by liquid scintillation counting (LSC) and AMS after radiochemical separation. AMS reaches an uncertainty < 10% at the 1 kBq gFe-1 level within less than 10 min measuring time. The background was < 3 Bq gFe-1, presently limited by the short measurement time. The new instrumental AMS method for analysing 55Fe from neutron capture production is reasonable and fast compared to other analytical methods. (author)

Additional details

Publishing Information

Journal Title
Journal of Radioanalytical and Nuclear Chemistry
Journal Volume
330
Journal Issue
3
Journal Page Range
p. 727-735
ISSN
0236-5731
CODEN
JRNCDM

Optional Information

Notes
34 refs.