Published 2006
| Version v1
Book
On the Traceability of Accuracy of Ultrasonic Flowmeter
Creators
- 1. Hokkaido University, Kita-ku, Sapporo (Japan)
Description
Time-of-flight ultrasonic flowmeters have been widely used these days in industry. It is however in suspicion if its high accuracy is traceable to the national standard. It was made clear why traceability cannot be guaranteed from a fluid mechanical point of view. The main reason is a difference of flow configuration between the flow standard and the measurement position on-site. The concept of 'Facility Factor' is introduced and it is concluded that the 'Profile Factor' is not sufficient for correcting the meter reading. It is discussed that measurement of velocity profile on-site is essentially required. (author)
Additional details
Publishing Information
- Publisher
- American Society of Mechanical Engineers - ASME
- Imprint Place
- New York (United States)
- Imprint Pagination
- 3 p.
Conference
- Title
- 14. International conference on nuclear engineering (ICONE 14)
- Dates
- 17-20 Jul 2006
- Place
- Miami - Florida (United States)
INIS
- Country of Publication
- United States
- Country of Input or Organization
- France
- INIS RN
- 39021580
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ACCURACY; CONFIGURATION; FLOWMETERS; FLUIDS; STANDARDS; TIME-OF-FLIGHT METHOD
- Descriptors DEC
- MEASURING INSTRUMENTS; METERS