  • 1. Department of Therapeutic Radiology, Yale University School of Medicine, 20 York Street, New Haven, Connecticut 06443 (United States)

Description

The methodology of brachytherapy source dosimetry with Gafchromic registered MD 55-2 film (ISP Technologies, Inc.) is examined with an emphasis on the nonlinearity of the optical density-dose relation within the dynamic dose range, the radial distance-dependent measurement uncertainty, and the format of data presentation. The specific source chosen for this study was a Checkmate trade mark sign (Cordis Corporation) intravascular brachytherapy system. The two-dimensional dose distribution around the source was characterized by a comprehensive analysis of measurement uncertainties. A comparative analysis of the dosimetric data from the vendor and from the scientific literature showed a substantial consistency of the information available for the Checkmate trade mark sign source. Our two-dimensional dosimetric data for the Checkmate trade mark sign source trains is presented in the form of measured along and away dose tables

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
33
Journal Issue
7
Journal Page Range
p. 2519-2524
ISSN
0094-2405
CODEN
MPHYA6

Optional Information

Notes