Published 1988 | Version v1
Report Restricted

Absolute calibration of a SPRED [Spectrometer Recording Extended Domain] EUV [extreme ultraviolet] spectrograph for use on the DIII-D tokamak

Description

We have performed an absolute intensity calibration of a SPRED multichannel EUV spectrograph using synchrotron radiation from the NBS SURF-II electron storage ring. The calibration procedure and results for both a survey grating (450 g/mm) and a high-resolution (2100 g/mm) grating are presented. The spectrograph is currently in use on the DIII-D tokamak with a tangential line-of-sight at the plasma midplane. Data is first acquired and processed by a microcomputer; the absolute line intensities are then sent to the DIII-D database for comparison with data from other diagnostics. Representative data from DIII-D plasma operations will be presented. 6 refs., 3 figs., 1 tab

Availability note (English)

Available from NTIS, PC A03/MF A01;1 as DE88011317.

Files

Restricted

The record is publicly accessible, but files are restricted to users with access.

Additional details

Publishing Information

Imprint Pagination
15 p.
Report number
UCRL--98278

Conference

Title
7. topical conference on high temperature plasma diagnostics.
Dates
13-17 Mar 1988.
Place
Napa, CA (USA).

Optional Information

Secondary number(s)
CONF-880364--29.