Radiological safety aspects of self-shielded medical cyclotron
Creators
- 1. Radiological Physics and Advisory Division, Bhabha Atomic Research Centre, Mumbai (India)
Description
Radiological safety around self-shielded medical cyclotrons, installed at two different nuclear medicine centers in the country has been assessed at the time of commissioning. An attempt has been made to measure gamma and neutron radiation levels inside the cyclotron vault/room. The mixed neutron field (thermal and fast neutron) has been measured using BF3 proportional counter and CR-39 foils. The gamma radiation levels have been monitored with the help of four dosimetric instruments: high pressurized ionization chamber, NaI(Tl) scintillation detector, diode based dosimeters and LiF:Mg,Cu,P Thermoluminescent (Tl) dosimeters. The gamma and neutron radiation levels inside the cyclotron vault/room were found in the range of 10.8 - 14.5 μSv/h and 0 - 16.8 μSv/h respectively. (author)
Additional details
Publishing Information
- Journal Title
- Radiation Protection and Environment
- Journal Volume
- 28
- Journal Issue
- 1-4
- Journal Page Range
- p. 144-146
- CODEN
- RPREFM
Conference
- Title
- 27. IARP national conference on occupational and environmental radiation protection
- Acronym
- IARPNC 2005
- Dates
- 23-25 Nov 2005
- Place
- Mumbai (India)
INIS
- Country of Publication
- India
- Country of Input or Organization
- India
- INIS RN
- 37111237
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CYCLOTRONS; DOSE RATES; GAMMA RADIATION; NEUTRON FLUX; NUCLEAR MEDICINE; RADIATION MONITORING; RADIATION PROTECTION; SELF-SHIELDING; THERMOLUMINESCENT DOSEMETERS
- Descriptors DEC
- ACCELERATORS; CYCLIC ACCELERATORS; DOSEMETERS; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; LUMINESCENT DOSEMETERS; MEASURING INSTRUMENTS; MEDICINE; MONITORING; RADIATION FLUX; RADIATIONS
Optional Information
- Notes