Published January 2000 | Version v1
Journal article

Amperometric detection of unithiol complexes of heavy metals in high-performance liquid chromatography

  • 1. Moskovskij Gosudarstvennyj Univ., Khimicheskij Fakul'tet, Moscow (Russian Federation)

Description

The feasibility is studied of Cd(2) indirect determination as its unithiol complexes by amperometric detection under static and HPLC conditions. Factors affecting the chromatographic separation and amperometric detection of metal complexes of unithiol were studied. Two designs of flow electrochemical cells were compared. The best sensitivity was attained for an amperometric detector with wall-jet flow cell and a graphic indicator electrode. The detection limits were 0.1 μg/ml. The chromatographic determination of heavy metals in a sample of waste water was carried out using the amperometric detector

Abstract (Russian)

Изучена возможность косвенного амперометрического детектирования Cd(2) в виде комплексов с унитиолом в статических условиях и в условиях ВЭЖХ. Исследованы факторы, влияющие на хроматографическиое разделение и амперометрическое детектирование комплексов металлов с унитиолом. Проведено сравнение двух конструкций проточных электрохимических ячеек. Максимальная чувствительность достигнута при использовании амперометрического детектора с проточной ячейкой типа стенка-сопло с индикаторным графитовым электродом. Пределы обнаружения для Cd(2) составляют 0,1 мкг/мл. Проведено хроматографическое определдение содержания тяжелых металлов в образце сточной воды с использованием амперометрического детектора

Additional details

Additional titles

Original title (Russian)
Амперометрическое детектирование комплексов тяжелых металлов с унитиолом в высокоэффективной жидкостной хроматограпхии

Publishing Information

Journal Title
Zhurnal Analiticheskoj Khimii
Journal Volume
55
Journal Issue
1
Journal Page Range
p. 59-65
ISSN
0044-4502
CODEN
ZAKHA8

Optional Information

Notes
8 refs., 5 tabs.