Published 1996 | Version v1
Journal article

Cryogenic aspects of the operation of a superconducting cyclotron-based neutron therapy facility

  • 1. Wayne State Univ., Detroit, MI (United States)
  • 2. Michigan State Univ., East Lansing, MI (United States)
  • 3. MedCyc Corporation, Haslett (United States)

Description

The Harper Hospital superconducting cyclotron, which is used for neutron radiation therapy, is a unique device. It is the first superconducting cyclotron to be installed in a hospital. The novel magnet cryostat can be rotated through 360 deg C without spilling liquid and whilst remaining vented to a low pressure line for collection of the boil-off gas. The mode of operation of the cryogenic magnet is described in detail. Some of the problems associated with the cryogenic nature of the cyclotron including those problems encountered in operating a helium liquefaction system in a hospital are discussed. At the present time the magnet is kept cold by filling the cryostat with approximately 75 L of liquid helium each day before patient treatments begin. This is a time-consuming process. The possibility of modifying the helium gas recovery and liquefaction system so that a continuous liquid helium supply cold be delivered to the magnet cryostat is discussed. (author)

Additional details

Additional titles

Original title (French)
Aspect cryogenique du fonctionnement d'un cyclotron supraconducteur utilise en neutrontherapie

Publishing Information

Journal Title
Bulletin du Cancer. Radiotherapie
Journal Volume
83
Series
4 refs.
Journal Page Range
p. 146-152.
ISSN
0924-4212
CODEN
BCRAEE

Conference

Title
Meeting of the European heavy particle therapy group.
Dates
4-6 May 1995.
Place
Orleans (France).