Published 2003 | Version v1
Miscellaneous

γ ray detection efficiency of BC-501A liquid scintillator

  • 1. China Academy of Engineering Physics, Mianyang (China). Inst. of Nuclear Physics and Chemistry

Description

The γ-ray detection efficiency of BC-501A detector was calculated using MARTHA M-C Code. The calculated results are coincident with the measured. The affection of aluminum container on detection efficiency increases with γ energy. For γ energy of 5 MeV, the ratio of detection efficiencies with and without the container is 1.35. The variance of the efficiencies with detection distances was studied. The accuracy of the calculated response functions was tested through handling the measured recoil electron spectrum from 24Na γ source

Part of:
Proceedings of the 11th national conference on nuclear electronics and nuclear detection technology

Additional details

Publishing Information

Imprint Title
Proceedings of the 11th national conference on nuclear electronics and nuclear detection technology
Imprint Pagination
598 p.
Journal Page Range
p. 480-483

Conference

Title
11. national conference on nuclear electronics and nuclear detection technology
Dates
1-6 Dec 2002
Place
Xiamen (China)

INIS

Country of Publication
China
Country of Input or Organization
China
INIS RN
35048918
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference, Non-conventional Literature
Descriptors DEI
ACCURACY; CONTAINERS; DISTANCE; EFFICIENCY; ENERGY DEPENDENCE; GAMMA DETECTION; LIQUID SCINTILLATION DETECTORS; M CODES; MONTE CARLO METHOD; RESPONSE FUNCTIONS
Descriptors DEC
CALCULATION METHODS; COMPUTER CODES; DETECTION; FUNCTIONS; MEASURING INSTRUMENTS; RADIATION DETECTION; RADIATION DETECTORS; SCINTILLATION COUNTERS