Published November 2016 | Version v1
Journal article

Calibration of a two-color soft x-ray diagnostic for electron temperature measurement

  • 1. University of Wisconsin - Madison, Madison, Wisconsin 53703 (United States)
  • 2. Consorzio RFX, Padova (Italy)
  • 3. Pierce College Fort Steilacoom, Lakewood, Washington 98498 (United States)

Description

The two-color soft x-ray (SXR) tomography diagnostic on the Madison Symmetric Torus is capable of making electron temperature measurements via the double-filter technique; however, there has been a 15% systematic discrepancy between the SXR double-filter (SXRDF) temperature and Thomson scattering (TS) temperature. Here we discuss calibration of the Be filters used in the SXRDF measurement using empirical measurements of the transmission function versus energy at the BESSY II electron storage ring, electron microprobe analysis of filter contaminants, and measurement of the effective density. The calibration does not account for the TS and SXRDF discrepancy, and evidence from experiments indicates that this discrepancy is due to physics missing from the SXRDF analysis rather than instrumentation effects.

Additional details

Identifiers

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
87
Journal Issue
11
Journal Page Range
vp.
ISSN
0034-6748
CODEN
RSINAK

Optional Information

Notes
(c) 2016 Author(s)