Published October 2006 | Version v1
Journal article

Standardization of 99mTc

  • 1. National Institute of R and D for Physics and Nuclear Engineering 'Horia Hulubei', IFIN-HH, P.O. Box MG-6, Bucharest, RO-77125 (Romania)

Description

The radioactivity of 99mTc was standardized by the 4πPC-γ coincidence method with two different modes. One is using coincidences between (119.5-142.6) keV conversion electrons and K X-rays, and the other is coincidences between the 2.13 keV conversion electrons and 140.5 keV γ-rays. The background of the K X-ray peak and the sensitivity of the proportional counter (PC) to 140 keV γ-rays were the main sources of uncertainties in the first case and low detection efficiency for conversion electrons in the second case. General coincidence equations were written, with specific forms, for the three measurement variants, including literature variant. Comparison with the ionization chamber calibration is reported

Additional details

Identifiers

DOI
10.1016/j.apradiso.2006.02.026;
PII
S0969-8043(06)00072-8;

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
64
Journal Issue
10-11
Journal Page Range
p. 1234-1237
ISSN
0969-8043
CODEN
ARISEF

Conference

Title
15. international conference on radionuclide metrology and its applications
Acronym
ICRM 2005
Dates
5-9 Sep 2005
Place
Oxford (United Kingdom)

Optional Information

Copyright
Copyright (c) 2006 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.