Evaluation of the retrospective ECG-gated helical scan using half-second multi-slice CT. Motion phantom study for volumetry

Description

ECG synchronized technique on multi-slice CT provide the thinner (less 2 mm slice thickness) and faster (0.5 sec/rotation) scan than that of the single detector CT and can acquire the coverage of the entire heart volume within one breath-hold. However, temporal resolution of multi-slice CT is insufficient on practical range of heart rate. The purpose of this study was to evaluate the accuracy of volumetry on cardiac function measurement in retrospective ECG-gated helical scan. We discussed the influence of the degradation of image quality and limitation of the heart rate in cardiac function measurement (volumetry) using motion phantom. (author)