Published March 1, 1993 | Version v1
Journal article

A novel charge sensitive preamplifier without the feedback resistor

  • 1. Politecnico di Milano, Dipt. di Elettronica e Informazione (Italy)
  • 2. Brookhaven National Lab., Upton, NY (United States)

Description

A novel charge sensitive preamplifier which has no resistor in parallel with the feedback capacitor is presented. No external device or circuit is required to discharge the feedback capacitor. The detector leakage and signal current flows away through the gate of the input JFET which works with its gate to source junction slightly forward biased. The DC stabilization of the preamplifier is accomplished by an additional feedback loop, which permits to equalize the current flowing through the forward biased gate to source junction and the current coming from the detector. An equivalent noise charge of less than 20 electrons rms has been measured at room temperature by using an input JFET with a transconductance to gate capacitance ratio of 4 mS/5.4 pF. (orig.)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A
Journal Volume
326
Journal Issue
1/2
Journal Page Range
p. 71-76.
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
New developments on radiation detectors.
Acronym
6. European symposium on semiconductor detectors
Dates
24-26 Feb 1992.
Place
Milan (Italy).

Optional Information

Contract/Grant/Project number
Contract DE-AC02-76CH00016