Published November 1, 2020 | Version v1
Journal article

Slow axis displacement correction for stripe artefact removal in optical coherence angiography

  • 1. Institute of Applied Physics of the RAS, Ulyanova St., 46, Nizhny Novgorod, 603950 (Russian Federation)

Description

The method for displacement correction along the slow axis of optical coherence tomography (OCT) data volumes is presented. The method is based on the recursive replacement of the next displaced OCT measurements by the weighted summation of itself and the previous OCT measurement in the slow axis dimension already corrected for the displacement. The values of the appropriate weight multipliers were defined from the local correlation of the two measurements. The proposed method was used as a preprocessing step for optical coherence angiography processing of the OCT data. A substantial reduction of the motion-induced stripe artifact was demonstrated. (letter)

Availability note (English)

Available from http://dx.doi.org/10.1088/1612-202X/ab96ca

Additional details

Identifiers

Publishing Information

Journal Title
Laser Physics Letters (Internet)
Journal Volume
17
Journal Issue
11
Journal Page Range
[3 p.]
ISSN
1612-202X

INIS

Country of Publication
Germany
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
53029992
Subject category
S71: CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSICS;
Descriptors DEI
BIOMEDICAL RADIOGRAPHY; BLOOD VESSELS; CORRECTIONS; TOMOGRAPHY
Descriptors DEC
BODY; CARDIOVASCULAR SYSTEM; DIAGNOSTIC TECHNIQUES; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIOLOGY