Published November 1, 2020
| Version v1
Journal article
Slow axis displacement correction for stripe artefact removal in optical coherence angiography
Creators
- 1. Institute of Applied Physics of the RAS, Ulyanova St., 46, Nizhny Novgorod, 603950 (Russian Federation)
Description
The method for displacement correction along the slow axis of optical coherence tomography (OCT) data volumes is presented. The method is based on the recursive replacement of the next displaced OCT measurements by the weighted summation of itself and the previous OCT measurement in the slow axis dimension already corrected for the displacement. The values of the appropriate weight multipliers were defined from the local correlation of the two measurements. The proposed method was used as a preprocessing step for optical coherence angiography processing of the OCT data. A substantial reduction of the motion-induced stripe artifact was demonstrated. (letter)
Availability note (English)
Available from http://dx.doi.org/10.1088/1612-202X/ab96caAdditional details
Identifiers
Publishing Information
- Journal Title
- Laser Physics Letters (Internet)
- Journal Volume
- 17
- Journal Issue
- 11
- Journal Page Range
- [3 p.]
- ISSN
- 1612-202X
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 53029992
- Subject category
- S71: CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSICS;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; BLOOD VESSELS; CORRECTIONS; TOMOGRAPHY
- Descriptors DEC
- BODY; CARDIOVASCULAR SYSTEM; DIAGNOSTIC TECHNIQUES; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIOLOGY