The effect of off-axis quality changes on zero area TAR for megavoltage beams
- 1. Northwestern Memorial Hospital, Chicago, IL (USA)
- 2. Northwestern Univ., Chicago, IL (USA). Dept. of Radiology
Description
In this work an analytical technique was employed to extract the primary component of dose at off-axis points from broad-beam water phantom data for a Clinac-18, Clinac-6(100), Clinac-4(Pb filter) and Philip's cobalt unit. The methodology presented here does not require separate off-axis HVL measurements and guarantees the accurate description of the data within the radiation field. (author)
Additional details
Publishing Information
- Journal Title
- Phys. Med. Biol.
- Journal Volume
- 30
- Journal Issue
- 6
- Series
- Phys. Med. Biol.
- Journal Page Range
- 589-595
- ISSN
- 0031-9155
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 16064043
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- BEAM PROFILES; LINEAR ACCELERATORS; MATHEMATICAL MODELS; PHANTOMS; PHOTON BEAMS; PLANNING; RADIATION QUALITY; RADIOTHERAPY; SPATIAL DOSE DISTRIBUTIONS; TISSUES
- Descriptors DEC
- ACCELERATORS; BEAMS; BODY; DISTRIBUTION; MEDICINE; MOCKUP; RADIATION DOSE DISTRIBUTIONS; SPATIAL DISTRIBUTION; STRUCTURAL MODELS; THERAPY