The effect of off-axis quality changes on zero area TAR for megavoltage beams

  • 1. Northwestern Memorial Hospital, Chicago, IL (USA)
  • 2. Northwestern Univ., Chicago, IL (USA). Dept. of Radiology

Description

In this work an analytical technique was employed to extract the primary component of dose at off-axis points from broad-beam water phantom data for a Clinac-18, Clinac-6(100), Clinac-4(Pb filter) and Philip's cobalt unit. The methodology presented here does not require separate off-axis HVL measurements and guarantees the accurate description of the data within the radiation field. (author)

Additional details

Publishing Information

Journal Title
Phys. Med. Biol.
Journal Volume
30
Journal Issue
6
Series
Phys. Med. Biol.
Journal Page Range
589-595