Published January 19, 2007 | Version v1
Journal article

Design of a High Resolution and High Flux Beam line for VUV Angle-Resolved Photoemission at UVSOR-II

  • 1. UVSOR Facility, Institute for Molecular Science, Okazaki 444-8585 (Japan)

Description

A high-energy-resolution angle-resolved photoemission beamline in the vacuum-ultraviolet (VUV) region has been designed for a 750 MeV synchrotron light source UVSOR-II. The beamline equips an APPLE-II-type undulator with the horizontally/vertically linear and right/left circular polarizations, a modified Wadsworth-type monochromator and a high-resolution photoelectron analyzer. The monochromator covers the photon energy range of 6 - 40 eV. The energy resolution (hv/Δhv) and the photon flux on samples are expected to be 2 x 104 and 1012 photons/sec at 10 eV, 4 x 104 and 5 x 1011 photons/sec at 20 eV, and 6 x 104 and 1011 photons/sec at 40 eV, respectively. The beamline provides the high-resolution angle-resolved photoemission spectroscopy less than 1 meV in the whole VUV energy range

Additional details

Identifiers

Publishing Information

Journal Title
AIP Conference Proceedings
Journal Volume
879
Journal Issue
1
Journal Page Range
p. 527-530
ISSN
0094-243X
CODEN
APCPCS

Conference

Title
9. international conference on synchrotron radiation instrumentation
Dates
28 May - 2 Jun 2006
Place
Daegu (Korea, Republic of)

Optional Information

Notes
(c) 2007 American Institute of Physics