The response of a spherical tissue-equivalent proportional counter to different heavy ions having similar velocities
Creators
- 1. Department of Environmental and Radiological Health Sciences, Colorado State University, Fort Collins, CO 80523 (United States)
- 2. Life Sciences Division, Lawrence Berkeley National Laboratory, Berkeley, CA 94720 (United States)
- 3. Department of Accelerator Physics and Engineering, National Institute of Radiological Sciences, 9-1 Anagawa 4-Chome, Inage-ku, Chiba 263 (Japan)
Description
A tissue-equivalent proportional counter (TEPC) has been used as a dosimeter in mixed radiation fields. Since it does not measure LET directly, the response function must be characterized in order to estimate quality factor and thus equivalent dose for the incident radiation. The objectives of this study were to measure the response of a spherical TEPC for different high-energy heavy ions (HZE) having similar velocity and to determine how quality factors can be determined. Data were obtained at the HIMAC heavy ion accelerator for 4He and 12C at 220+/-5MeV/nucleon(β=0.59) and 12C, 16O, 28Si and 56Fe at 376+/-15MeV/nucleon(β=0.70). A particle spectrometer recorded the charge and position of each incident beam particle. Events with low energy deposition were observed for particles that passed through the wall of the TEPC but not through the sensitive volume. The frequency averaged lineal energy, y-barf, was always less than the LET of the incident particles. The dose averaged lineal energy, y-bar D, was approximately equal to LET for particles with LET greater than 10keV/μm, whereas y-barD was larger than LET for the lighter particles with lower LET. Part of this effect is due to detector resolution and energy straggling that increases the variance of the response function. Although the TEPC is not a LET spectrometer, it can provide real time measurements of dose and provide estimates of quality factors for HZE particles using averaged values of lineal energy
Additional details
Identifiers
- DOI
- 10.1016/j.radmeas.2006.01.003;
- PII
- S1350-4487(06)00153-3;
Publishing Information
- Journal Title
- Radiation Measurements
- Journal Volume
- 41
- Journal Issue
- 9-10
- Journal Page Range
- p. 1227-1234
- ISSN
- 1350-4487
- CODEN
- RMEAEP
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 38013065
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- ASTATINE 220; CARBON 12; DOSEMETERS; HEAVY ION ACCELERATORS; HEAVY IONS; HELIUM 4; IRON 56; LET; OXYGEN 16; PARTICLES; PROPORTIONAL COUNTERS; QUALITY FACTOR; RADIATION DOSES; RESPONSE FUNCTIONS; SILICON 28; SPECTROMETERS; TIME MEASUREMENT
- Descriptors DEC
- ACCELERATORS; ALPHA DECAY RADIOISOTOPES; ASTATINE ISOTOPES; BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; CARBON ISOTOPES; CHARGED PARTICLES; DIMENSIONLESS NUMBERS; DOSES; ENERGY TRANSFER; EVEN-EVEN NUCLEI; FUNCTIONS; HEAVY NUCLEI; HELIUM ISOTOPES; INTERMEDIATE MASS NUCLEI; IONS; IRON ISOTOPES; ISOTOPES; LIGHT NUCLEI; MEASURING INSTRUMENTS; MINUTES LIVING RADIOISOTOPES; NUCLEI; ODD-ODD NUCLEI; OXYGEN ISOTOPES; RADIATION DETECTORS; RADIOISOTOPES; SILICON ISOTOPES; STABLE ISOTOPES
Optional Information
- Copyright
- Copyright (c) 2006 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.