Published May 1981 | Version v1
Journal article

BRH test pattern for the evaluation of gamma-camera performance

  • 1. Bureau of Radiological Health, Rockville, MD

Description

A transmission test pattern has been developed for the simultaneous evaluation of intrinsic resolution, field uniformity, and spatial distortion of gamma cameras of any model and age. Holes 2.5 mm in diameter are drilled with high precision through a 3.2-mm thick lead-alloy plate, forming an orthogonal array of holes. The performance of gamma cameras has been investigated with the gamma rays of Tc-99m and the low-energy photons from Xe-133 and Tl-201. Images of the test pattern provide a simple, rapid, and sensitive method for gamma camera quality control, acceptance testing of new instruments, and checking of a camera's operating and maintenance procedures

Additional details

Publishing Information

Journal Title
J. Nucl. Med.
Journal Volume
22
Journal Issue
5
Series
J. Nucl. Med.
Journal Page Range
468-470
ISSN