Estimation of effective dose of dental X-ray devices
Creators
- 1. Changzhou CDC, No. 203, Taishan Road, Changzhou, Jiangsu 213022 (China)
Description
This study aims to estimate the effective doses of dental X-ray devices under common scanning protocols. After putting TLDs in the Alderson Radiation Therapy Phantom, we exposed the phantom under common scanning protocols of three dental X-ray devices, namely CBCT, dental panoramic machine and intraoral round cone device. Then effective doses were calculated using the measured absorbed doses of organs and tissues. Tissue weighting factors recommended by the ICRP were adopted in the calculation. Effective doses under common scanning protocols of three Dental X-ray devices were obtained. The effective dose of dental CT was 0.20 mSv, and that of dental panoramic machine and intraoral radiography were 0.013 and 0.0050 mSv, respectively. The tissue absorbed doses of dental CT scan were 0.63 mGy of brain, 7.7 mGy of salivary glands, 8.7 mGy of thyroid and 4.0 mGy of the lens of the eye. The tissue absorbed doses from dental panoramic machine are 0.62 mGy of salivary glands and 0.25 mGy of thyroid. And finally the tissue absorbed dose of intraoral radiography was 0.80 mGy of salivary gland. Among the three dental X-ray devices studied, dental CBCT scan can cause much higher effective dose than the other two. Brain, salivary glands, thyroid and the lens of the eye are tissues receiving relatively higher absorbed doses. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1093/rpd/ncy159Additional details
Identifiers
- DOI
- 10.1093/rpd/ncy159;
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 183
- Journal Issue
- 4
- Journal Page Range
- p. 418-422
- ISSN
- 0144-8420
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- France
- INIS RN
- 50047234
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ABSORBED RADIATION DOSES; ANIMAL TISSUES; BRAIN; COMPUTERIZED TOMOGRAPHY; CRYSTALLINE LENS; EFFECTIVE RADIATION DOSES; ICRP; PHANTOMS; RADIOTHERAPY; SALIVARY GLANDS; THYROID
- Descriptors DEC
- BODY; CENTRAL NERVOUS SYSTEM; DIAGNOSTIC TECHNIQUES; DOSES; ENDOCRINE GLANDS; EYES; FACE; GLANDS; HEAD; INTERNATIONAL ORGANIZATIONS; MEDICINE; MOCKUP; NERVOUS SYSTEM; NUCLEAR MEDICINE; ORGANS; RADIATION DOSES; RADIOLOGY; SENSE ORGANS; STRUCTURAL MODELS; THERAPY; TOMOGRAPHY
Optional Information
- Notes
- 23 refs.