Published April 1, 1983 | Version v1
Journal article

A large-area xenon gas scintillation proportional counters (GSPC) with timing information for the detection of low energy X-rays

  • 1. Eidgenoessische Technische Hochschule, Zurich (Switzerland). Lab. fuer Hochenergiephysik

Description

A large area xenon gas scintillation proportional counter has been developed for the study of low energy X-rays. A time resolution of 40 ns (fwhm) at 5.5 keV was obtained by producing secondary scintillation light in the drift space. A simple model of the processes leading to secondary light production describes the measured time distributions and their energy dependence. The energy resolution at 5.9 keV was 8.8% (fwhm) for collimated X-rays. For a uniform irradiation over the sensitive area of 314 cm2, a resolution of 9.5% is obtained after correcting for the pulse height dependence on the position of X-ray absorption. The reduction of pulse height due to small H2O and O2 contaminations was measured. (orig.)

Additional details

Publishing Information

Journal Title
Nucl. Instrum. Methods Phys. Res.
Journal Volume
207
Journal Issue
3
Series
Nucl. Instrum. Methods Phys. Res.
Journal Page Range
429-435
ISSN
0029-554X

Optional Information

Notes
With 9 figs.