Published October 2012 | Version v1
Journal article

The PTB primary standard for the absorbed-dose to water for I-125 interstitial brachytherapy sources

Creators

  • 1. Physikalisch-Technische Bundesanstalt (PTB), Braunschweig, (Germany)

Description

The German national metrology institute (PTB) developed a primary standard in terms of absorbed-dose to water Dw for low-energy interstitial brachytherapy sources, which is based on an extrapolation chamber in a phantom of water-equivalent material. The method to determine Dw from extrapolation chamber measurements has been newly developed and is already described in the literature. With the chamber the absorbed-dose at 30 cm distance from the source is measured and the quantity is converted into the desired quantity, the absorbed-dose to water measured at 1 cm distance perpendicular to the source axis. In this paper, a synthesis of the work done within the EMRP Project: 'TP2.JRP6: Increasing Cancer Treatment Efficacy Using 3D Brachytherapy' is given and the final results and the final uncertainty budget are presented. Furthermore, an experimentally determined dose-rate constant for this seed type (BEBIG Symmetra I25.S16) is given based on the measurement of four different instances. (author)

Availability note (English)

Available from doi: http://dx.doi.org/10.1088/0026-1394/49/5/S198

Additional details

Identifiers

Publishing Information

Journal Title
Metrologia
Journal Volume
49
Journal Issue
no.5
Journal Page Range
p. S198-S202
ISSN
0026-1394
CODEN
MTRGAU

Conference

Title
Conference on Advanced Metrology for Cancer Therapy 2011
Dates
29 Nov - 1 Dec 2011
Place
Braunschweig (Germany)

Optional Information

Notes
14 refs.