Published September 1981
| Version v1
Journal article
Design for a multiple target system for a medical cyclotron
Creators
- 1. Biophysics Laboratory, Memorial Sloan-Kettering Cancer Center, New York, New York 10021
Description
A novel target system has been designed for a compact cyclotron. The system permits one of three or more targets to be selected remotely for bombardment, permits all target and chemical operations to be controlled remotely, causes no change in beam energy or beam current, occupies a limited space, and is simple to construct and maintain. The targets are mounted to a flexible bellows-multiple port manifold assembly, supported on a fan-shaped turntable. A motor moves the system to align the target to be bombarded with the main beam axis. Designs for two subassemblies are also presented. These are a retractable target chamber for water and a remotely controlled variable collimator
Additional details
Publishing Information
- Journal Title
- Med. Phys.
- Journal Volume
- 8
- Journal Issue
- 5
- Series
- Med. Phys.
- Journal Page Range
- 703-705
- ISSN
- 0094-2405
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 13665213
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES;
- Descriptors DEI
- CYCLOTRONS; PARTICLE BEAMS; RADIOISOTOPES; SPECIFICATIONS; TARGETS
- Descriptors DEC
- ACCELERATORS; BEAMS; CYCLIC ACCELERATORS; ISOTOPES