Published February 1, 1981 | Version v1
Journal article

Optimum weighted double sampling technique for pulse amplitude measurement

  • 1. Magyar Tudomanyos Akademia Atommag Kutato Intezete, Debrecen

Description

According to the presented technique the pulse amplitude is measured as the difference between the attenuated instantaneous noise amplitude preceding the pulse and the maximum of the pulse. Baseline fluctuation for high counting rates is avoided by the introduction of an appropriate time variant filter. Theoretical calculations for some pulse shaping circuits are given and the realization of the technique in question with experimental results is shown. (orig.)

Additional details

Publishing Information

Journal Title
Nucl. Instrum. Methods
Journal Volume
179
Journal Issue
3
Series
Nucl. Instrum. Methods.
Journal Page Range
605-613
ISSN
0029-554X

INIS

Country of Publication
Netherlands
Country of Input or Organization
Netherlands
INIS RN
12614245
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
AMPLITUDES; COUNTING RATES; ELECTRONIC EQUIPMENT; FLUCTUATIONS; PULSE SHAPERS; PULSES; SAMPLING
Descriptors DEC
ELECTRONIC CIRCUITS; PULSE CIRCUITS; VARIATIONS