Published February 1975
| Version v1
Journal article
Gated baseline restorer with adjustable asymmetry
Description
A baseline restorer (BLR) for pulse amplifiers is described in which current to the memory capacitor can be gated off during a pulse. This technique reduces the BLR recovery time without introducing artifacts when a short-period undershoot accompanies in input signal. The gating and reset thresholds are adjustable, as is the BLR symmetry. Tests with a pulse generator show that the baseline shift with 5V pulses at duty factors up to 75 percent is less than 1mV. When the signal source is a radiation detector, the gated BLR is usable at duty factors up to 90 percent. (U.S.)
Additional details
Identifiers
Publishing Information
- Journal Title
- IEEE Transactions on Nuclear Science
- Journal Volume
- 22
- Journal Issue
- 1
- Series
- IEEE (Inst. Electr. Electron. Eng.) Trans. Nucl. Sci.
- Journal Page Range
- 463-466
- ISSN
- 0018-9499
Conference
- Title
- 21. nuclear science symposium; 14. scintillation and semiconductor counter symposium; 6. nuclear power systems symposium.
- Dates
- 11 Dec 1974.
- Place
- Washington, DC.
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 6201954
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ELECTRONIC CIRCUITS; GAIN; GAMMA SPECTROMETERS; LI-DRIFTED GE DETECTORS; PERFORMANCE; PULSE AMPLIFIERS; STABILITY
- Descriptors DEC
- AMPLIFIERS; GE SEMICONDUCTOR DETECTORS; LI-DRIFTED DETECTORS; MEASURING INSTRUMENTS; RADIATION DETECTORS; SEMICONDUCTOR DETECTORS; SPECTROMETERS
Optional Information
- Notes
- Updated automatically by Metadata and Full-Text Enrichment Agent