Published October 2004 | Version v1
Journal article

Error estimation and parameter dependence of the calculation of the fast ion distribution function, temperature, and density using data from the KF1 high energy neutral particle analyzer on Joint European Torus

  • 1. Helsinki University of Technology, Association EURATOM-TEKES, P.O. Box 2200, FIN-02015 HUT (Finland)
  • 2. Consorzio RFX, Associazione EURATOM-ENEA sulla Fusione, I-35127 Padova (Italy)
  • 3. Division of Fusion Plasma Physics, Alfven Laboratory, Royal Institute of Technology, Association EURATOM NFR, Stockholm (Sweden)
  • 4. Centre de Recherches en Physique des Plasmas, Association EURATOM Confederation Suisse, Ecole Polytechnique Federale de Lausanne, CRPP EPFL, CH-1015 Lausanne (Switzerland)

Description

Joint European Torus high energy neutral particle analyzer measures the flux of fast neutrals originating from the plasma core. From this data, the fast ion distribution function fifast, temperature Ti,perpendicularfast, and density nifast are derived using knowledge of various plasma parameters and of the cross section for the required atomic processes. In this article, a systematic sensitivity study of the effect of uncertainties in these quantities on the evaluation of the neutral particle analyzer fifast, Ti,perpendicularfast, and nifast is reported. The dominant parameter affecting nifast is the impurity confinement time and therefore a reasonable estimate of this quantity is necessary to reduce the uncertainties in nifast below 50%. On the other hand, Ti,perpendicularfast is much less sensitive and can certainly be provided with an accuracy of better than 10%

Additional details

Identifiers

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
75
Journal Issue
10
Journal Page Range
p. 3547-3549
ISSN
0034-6748
CODEN
RSINAK

Optional Information

Notes
(c) 2004 American Institute of Physics
Collaborations
JET-EFDA Contributors