Published June 2009 | Version v1
Journal article

A mathematical method to calculate efficiency of BF3 detectors

  • 1. Institute of Nuclear Physics and Chemistry, China Academy of Engineering Physics, Mianyang (China)

Description

In order to calculate absolute efficiency of the BF3 detector, MCNP/4C code is applied to calculate relative efficiency of the BF3 detector first, and then absolute efficiency is figured out through mathematical techniques. Finally an energy response curve of the BF3 detector for 1-20 MeV neutrons is derived. It turns out that efficiency of BF3 detector are relatively uniform for 2-16 MeV neutrons. (authors)

Additional details

Publishing Information

Journal Title
Nuclear Science and Techniques
Journal Volume
20
Journal Issue
3
Journal Page Range
p. 181-183
ISSN
1001-8042

Optional Information

Notes
3 figs., 14 refs.