Published July 2011 | Version v1
Journal article

Evaluation of image quality and radiation dose at prospective ECG-triggered axial 256-slice multi-detector CT in infants with congenital heart disease

  • 1. Guangdong Academy of Medical Sciences, Guangdong General Hospital, Department of Radiology, Guangzhou (China)
  • 2. Philips Healthcare, CT Clinical Science, Highland Heights, OH (United States)

Description

There are a limited number of reports on the technical and clinical feasibility of prospective electrocardiogram (ECG)-gated multi-detector computed tomography (MDCT) in infants with congenital heart disease (CHD). To evaluate image quality and radiation dose at weight-based low-dose prospectively gated 256-slice MDCT angiography in infants with CHD. From November 2009 to February 2010, 64 consecutive infants with CHD referred for pre-operative or post-operative CT were included. All were scanned on a 256-slice MDCT system utilizing a low-dose protocol (80 kVp and 60-120 mAs depending on weight: 60 mAs for ≤3 kg, 80 mAs for 3.1-6 kg, 100 mAs for 6.1-10 kg, 120 mAs for 10.1-15 kg). No serious adverse events were recorded. A total of 174 cardiac deformities, confirmed by surgery or heart catheterization, were studied. The sensitivity of MDCT for cardiac deformities was 97.1%; specificity, 99.4%; accuracy, 95.9%. The mean heart rate during scan was 136.7 ± 14.9/min (range, 91-160) with a corresponding heart rate variability of 2.8 ± 2.2/min (range, 0-8). Mean scan length was 115.3 ± 11.7 mm (range, 93.6-143.3). Mean volume CT dose index, mean dose-length product and effective dose were 2.1 ± 0.4 mGy (range, 1.5-2.8), 24.7 ± 5.9 mGy.cm (range, 14.7-35.8) and 1.6 ± 0.3 mSv (range, 1.1-2.5), respectively. Diagnostic-quality images were achieved in all cases. Satisfactory diagnostic quality for visualization of all/proximal/distal coronary artery segments was achieved in 88.4/98.8/80.0% of the scans. Low-dose prospectively gated axial 256-slice CT angiography is a valuable tool in the routine clinical evaluation of infants with CHD, providing a comprehensive three-dimensional evaluation of the cardiac anatomy, including the coronary arteries. (orig.)

Availability note (English)

Available from: http://dx.doi.org/10.1007/s00247-011-2079-2

Additional details

Identifiers

Publishing Information

Journal Title
Pediatric Radiology
Journal Volume
41
Journal Issue
7
Journal Page Range
p. 858-866
ISSN
0301-0449
CODEN
PDRYA5