Published December 2015 | Version v1
Journal article

Study of a new neutron dosimeter incorporating RPL detectors

  • 1. Institut Pluridisciplinaire Hubert Curien (IPHC), University of Strasbourg, CNRS-IN2P3-UMR 7178, 23 rue du Loess, BP 28, F-67037, Strasbourg, Cedex 2 (France)
  • 2. Faculté Poly-disciplinaire de Safi, Université Cadi Ayyad, Route Sidi Bouzid BP 4162, 46000, Safi (Morocco)

Description

We characterize a passive dosimeter incorporating RadioPhotoLuminescent (RPL) detectors capable of measuring both fast and thermal neutrons for personal dosimetry. These neutrons can be detected in a mixed n-γ field with appropriate converters. Monte Carlo simulations with MCNPX helped with the geometrical conception of the dosimeter and the choice of materials. The responses of the RPL dosimeter are linear in Hp(10) with detection limits of 1.83 mSv for fast neutrons and 0.19 mSv for thermal neutrons. The angular dependencies are satisfactory according to the ISO 21909 norm. A calibration factor of (9.5 ± 0.5) × 10−2 mSv cm2/RPL intensity was obtained for the fast neutrons of the IPHC's 241Am–Be calibrator. The factor was (9.7 ± 0.3) × 10−3 mSv cm2/RPL intensity for thermalized neutrons.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.radmeas.2015.05.004

Additional details

Identifiers

DOI
10.1016/j.radmeas.2015.05.004;
PII
S1350-4487(15)30022-6;

Publishing Information

Journal Title
Radiation Measurements
Journal Volume
83
Journal Page Range
p. 47-50
ISSN
1350-4487
CODEN
RMEAEP

Conference

Title
26. international conference on nuclear tracks in solids
Acronym
ICNTS-26
Dates
15-19 Sep 2014
Place
Kobe (Japan)

Optional Information

Copyright
Copyright (c) 2015 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.