Published May 2015 | Version v1
Miscellaneous

Ultra Low Level Tritium Analysis Method Using a Liquid Scintillation Counter

  • 1. Dongnam Institute of Radiological and Medical Sciences, Busan (Korea, Republic of)
  • 2. Inje University, Gimhae, (Korea, Republic of)

Description

To evaluate 3H concentration in the atmosphere more accurately compared to the conventional methods, the author of this paper intended to suggest more improved analytical methods and derived the elements which might occur during analysis or required improvements. The method suggested in this study is able to reduce the uncertainty and errors which may be existent in evaluating the 3H concentration of environmental sample s and thus will serve as the best solution in the technical and economic point of view. Liquid Scintillation Counter is the most widely used to analyze ultra-low level 3H by using CPM / DPM Counting Mode using external radiation source and Spectrum Plot Mode using internal radiation source. In CPM / DPM Counting Mode, multiple samples can be measured by single calibration despite its rather higher background whereas Spectrum Plot Mode requires more time and cost to analyze multiple samples despite its reliability to reduce the contribution of other radionuclides

Part of:
Proceedings of the KNS 2015 Spring Meeting

Additional details

Publishing Information

Publisher
KNS
Imprint Place
Daejeon (Korea, Republic of)
Imprint Title
Proceedings of the KNS 2015 spring meeting
Imprint Pagination
[1 CD-ROM]
Journal Page Range
[2 p.]

Conference

Title
2015 spring meeting of the KNS
Dates
6-8 May 2015
Place
Jeju (Korea, Republic of)

Optional Information

Notes
5 refs, 3 figs