Published March 2002 | Version v1
Journal article

Precision in Moessbauer Spectroscopy

  • 1. Institute for Analytical Instrumentation (Russian Federation)
  • 2. Federal Institute for Materials Research and Testing (BAM) (Germany)
  • 3. Radium Institute (Russian Federation)

Description

In this paper the factors which are responsible for precision and reliability of the experimental data are discussed. The detectors of different types are compared with respect to the precision of the final results. It is shown that the use of resonance detectors gives the best results in most cases of the Sn-119m Moessbauer spectroscopy. The estimation of spectral line distortion caused by a short distance between source and detector is presented. Different procedures of velocity scale calibration are discussed. It has to be underlined that Kuprijanov's method of the velocity scale calibration gives the most reliable and fast results.

Additional details

Identifiers

Publishing Information

Journal Title
Hyperfine Interactions
Journal Volume
139-140
Journal Issue
1-4
Journal Page Range
p. 679-684
ISSN
0304-3843
CODEN
HYINDN

Optional Information

Copyright
Copyright (c) 2002 Kluwer Academic Publishers