An accurate calibration method of the multileaf collimator valid for conformal and intensity modulated radiation treatments
- 1. Department of Radiotherapy, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX Utrecht (Netherlands)
Description
Because for IMRT treatments the required accuracy on leaf positioning is high, conventional calibration methods may not be appropriate. The aim of this study was to develop the tools for an accurate MLC calibration valid for conventional and IMRT treatments and to investigate the stability of the MLC. A strip test consisting of nine adjacent segments 2 cm wide, separated by 1 mm and exposed on Kodak X-Omat V films at Dmax depth, was used for detecting leaf-positioning errors. Dose profiles along the leaf-axis were taken for each leaf-pair. We measured the dose variation on each abutment to quantify the relative positioning error (RPE) and the absolute position of the abutment to quantify the absolute positioning error (APE). The accuracy of determining the APE and RPE was 0.15 and 0.04 mm, respectively. Using the RPE and the APE the MLC calibration parameters were calculated in order to obtain a flat profile on the abutment at the correct position. A conventionally calibrated Elekta MLC was re-calibrated using the strip test. The stability of the MLC and leaf-positioning reproducibility was investigated exposing films with 25 adjacent segments 1 cm wide during three months and measuring the standard deviation of the RPE values. A maximum shift over the three months of 0.27 mm was observed and the standard deviation of the RPE values was 0.11 mm
Availability note (English)
Available online at http://stacks.iop.org/0031-9155/49/2631/pmb4_12_011.pdf or at the Web site for the journal Physics in Medicine and Biology (ISSN 1361-6560) http://www.iop.org/Additional details
Identifiers
- URL
- http://stacks.iop.org/0031-9155/49/2631/pmb4_12_011.pdf; http://www.iop.org/;
- DOI
- 10.1088/0031-9155/49/12/011;
- PII
- S0031-9155(04)71864-3;
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 49
- Journal Issue
- 12
- Journal Page Range
- p. 2631-2643
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 35068926
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCURACY; CALIBRATION; COLLIMATORS; ERRORS; FILMS; POSITIONING; RADIOTHERAPY
- Descriptors DEC
- MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; THERAPY