TU-FG-209-07: Medical Physics 1.0 Versus Medical Physics 2.0: A Case Study

  • 1. University of Texas MD Anderson Cancer Center, Houston, TX (United States)
  • 2. Duke University Medical Center, Durham, NC (United States)

Description

Purpose: To illustrate how performance analytics can identify performance decrement in digital radiography systems. Methods: Subsequent to a radiologist's image quality complaint, four different advanced methods contributed to root cause analysis. Our system was a GE Revolution XQi digital radiography unit. Initially, we reviewed weekly GE Quality Assurance Procedures (QAP) results in a database dating from 2001. Next, we evaluated objective image quality metrics of individual PA Chest radiographs acquired. These images were anonymized, securely transferred, and analyzed by the Duke University Clinical Imaging Physics Group with software previously described1 and validated2. Third, we compared the exposure-dependent SNR2 (NEQ) of the unit with previously established confidence limits3. Finally, we explored our service database to reveal events that might affect detector performance. Results: QAP reported a decrease in CNR reflected in a significant increase in lung noise(Ln), mediastinum noise(Mn), and subdiaphragm-lung contrast(Slc) with a significant decrease in lung grey level(Lgl) after detector replacement. Most change occurred during week 1, before the QAP indicated one-half the ultimate decrease in CNR. After detector recalibration, QAP CNR improved, but was not restored to previous levels. Lgl and Slc were no longer significantly different from before, however Ln and Mn remained significantly different. Exposure-dependent SNR2 show the detector to be operating within limits in October 2006 but subsequently became miscalibrated sometime before acquisition of the 2011–2014 data. Service records revealed catastrophic failure of the Image Detection Controller that contained the 2007 calibration. Traditional metrics did not indicate that the system was performing outside of normal limits. Conclusion: Performance analytics are powerful tools whose proper application could allow early intervention in degraded system performance. The image-quality metrics appear to be highly sensitive to system performance and are reported with every acquisition rather than at arbitrary intervals. Confidence intervals may require customization for individual systems or detectors.

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
43
Journal Issue
6
Journal Page Range
vp.
ISSN
0094-2405
CODEN
MPHYA6

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
48102925
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY; S60: APPLIED LIFE SCIENCES;
Descriptors DEI
AGE ESTIMATION; BIOMEDICAL RADIOGRAPHY; COMPUTER CODES; EDUCATIONAL FACILITIES; IMAGES; METRICS; PERFORMANCE; QUALITY ASSURANCE
Descriptors DEC
DIAGNOSTIC TECHNIQUES; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY

Optional Information

Notes
(c) 2016 American Association of Physicists in Medicine