Quality control of scintillation cameras (planar and SPECT)
- 1. Department of Medical Physics, Atomic Energy Council, Sudan Academy of Sciences, Khartoum (Sudan)
Description
Regular quality control is one of the corner stones of nuclear medicine and a prerequisite for adequate diagnostic imaging. Many papers have been published on quality control of planar and SPECT imaging system up to now, however only minor attenuation has been given to the assessment of the performance of imaging systems. In this research we are going to discuss a comprehensive set of test procedures including regular quality control. Our purpose is to go through analysis of the methods and results then to test our hypothesis which state that there is strong relationship between regular and proper evaluation of quality control and the continuity of better medical services in nuclear medicine department. The selection of the tests is discussed and the tests are described, then results are presented. In addition action thresholds are proposed. The quality control tests can be applied to systems with either a moving detector or a moving image table, and to both detector with a large field of view and detectors with a small field of view. The tests presented on this research do not required special phantoms or sources other than those used for quality control of stationary gamma camera and SPECT. They can be applied for acceptance testing and for performance testing in a regular quality assurance program. The data has been evaluated based on me diso software in comparing with IAEA expert software and system specification within the reference values. Our final results confirm our hypothesis, there are some comments about the characteristics and performance of this system that being observed and solved, then a departmental protocol for routine quality control (Q.C) has being established.(Author)
Availability note (English)
Available from INIS in electronic form; Also available from Sudan Atomic Energy Commission, Khartoum (SD)Files
39093617.pdf
Files
(9.0 MB)
| Name | Size | Download all |
|---|---|---|
|
md5:a63e8ac9c0d19845b369394156d84176
|
9.0 MB | Preview Download |
Additional details
Publishing Information
- Imprint Place
- Khatoum (Sudan)
- Imprint Pagination
- 63 p.
- Report number
- INIS-SD--324
INIS
- Country of Publication
- Sudan
- Country of Input or Organization
- Sudan
- INIS RN
- 39093617
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Thesis
- Descriptors DEI
- DIAGNOSTIC TECHNIQUES; GAMMA CAMERAS; IMAGES; NUCLEAR MEDICINE; PERFORMANCE TESTING; PHANTOMS; QUALITY CONTROL; SCINTILLATION COUNTERS; SUDAN
- Descriptors DEC
- AFRICA; ARAB COUNTRIES; CAMERAS; CONTROL; DEVELOPING COUNTRIES; MEASURING INSTRUMENTS; MEDICINE; MOCKUP; RADIATION DETECTORS; STRUCTURAL MODELS; TESTING
Optional Information
- Notes
- 25 refs., 7 tabs., 19 figs.