Published May 29, 1995 | Version v1
Journal article

Performance of HIMAC

  • 1. Research Center for Nuclear Physics, Osaka University, 10-1, Mihogaoka, Ibaraki-shi, Osaka 567 (Japan)
  • 2. Research Center of Charged Particle Therapy, Chiba (Japan). Nat. Inst. Radiol. Sci.

Description

The NIRS heavy-ion two-synchrotron medical facility, HIMAC, was approved in the 1987 fiscal year and clinical trials were started in late June 1994 as previously arranged. The operation experiences show that the entire HIMAC facility can work well with high stability and excellent reproducibility. For example, the intensity of beam extracted slowly from the rings can be reduced as low as 500 particles per pulse because stable and reproducible acceleration can be achieved in the rings in spite of no beam feedback. This performance enables direct counting of the beam and its fragments in the preparatory experiments toward clinical treatments and provides a promising basis for acceleration and storage of radioactive beams toward simultaneous treatment and diagnosis in future. ((orig.))

Additional details

Publishing Information

Journal Title
Nuclear Physics. A
Journal Volume
588
Journal Issue
1
Journal Page Range
p. 229C-234C.
ISSN
0375-9474
CODEN
NUPABL

Conference

Title
5. international symposium on physics of unstable nuclei.
Dates
31 Oct - 3 Nov 1994.
Place
Niigata (Japan).

INIS