Published May 2014 | Version v1
Journal article

Standardization of 18F and its use for the Romanian PET metrological traceability chain assurance

  • 1. Horia Hulubei National Institute for R and D in Physics and Nuclear Engineering, IFIN-HH, 30 Reactorului St., Magurele, jud. Ilfov, POB MG-6, RO 0-77125, Bucharest (Romania)
  • 2. Monrol Europe srl., Pantelimon (Romania)

Description

This paper presents the work performed at IFIN-HH and at the users' sites, aimed to establish the 18FDG Romanian PET traceability chain. It summarizes the operations: (i) Absolute standardization of a 18F solution, by the 4πβ(PC)-γ coincidence method; (ii) Control of gamma-ray impurity content and measurement of the activity by gamma ray spectrometry; (iii) Measurement of the half life, for detection of some short/long life positron emitter impurities; (iv) Calibration of the CENTRONIC IG12/20A ionization chamber; (v) Calibration of commercial radionuclide calibrators. - Highlights: • Presents establishment of the 18F FDG Romanian PET traceability chain. • Shows work at IFIN-HH and at the users' site. • An 18F solution was standardized by the 4πβ(PC)-γ coincidence method. • Includes measurement of half-life for detection of positron emitting impurities. • CENTRONIC IG12/20A ionization chamber and radionuclide calibrators calibrated

Availability note (English)

Available from http://dx.doi.org/10.1016/j.apradiso.2013.11.047

Additional details

Identifiers

DOI
10.1016/j.apradiso.2013.11.047;
PII
S0969-8043(13)00493-4;

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
87
Journal Page Range
p. 14-18
ISSN
0969-8043
CODEN
ARISEF

Conference

Title
19. international conference on radionuclide metrology and its applications
Dates
17-21 Jun 2013
Place
Antwerp (Belgium)

Optional Information

Copyright
Copyright (c) 2013 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.